2100Q Portable Turbidimeter Hach
2100Q Portable Turbidimeter Hach
The 2100Q portable turbidimeter is compliant with USEPA Method 180.1 design criteria.
Easy on-screen assisted calibration and verification
Simple data transfer
Accurate for rapidly settling samples
Convenient data logging
Optical system for precision in the field
Specifications :
Measurement Method   Ratio turbidimetric determination using a primary
                                 nephelometric light scatter signal (90°) to the transmitted 
                                 light scatter signal.
Regulatory                 2100Q: Meets EPA Method 180.1
                                2100Q is: Meets ISO 7027 Light Source
                                2100Q: Tungsten filament lamp 
                                2100Q is: Light-emitting diode (LED) @ 860 nm 
Range                        0 to 1000 NTU (FNU)
Accuracy                   ±2% of reading plus stray light from 0 to 1000 NTU
Repeatability              ±1% of reading, or 0.01 NTU (FNU), whichever is greater
Resolution                 0.01 NTU on lowest range
Stray Light                <0.02 NTU (FNU)
Signal Averaging        Selectable on/off
Detector                   Silicon photovoltaic
Reading Modes (user selectable)
Normal (Push to Read) Signal Averaging
                                 Rapidly Settling Turbidity
Data Logger               500 records
Power Requirement      110-230 Vac, 50/60 Hz (with Power or USB+Power Module)
                                 4 AA alkaline batteries
                                 Rechargeable NiMH (for use with USB+Power Module
Operating Conditions   Temperature: 0 to 50°C (32 to 122°F)
                                 Relative Humidity: 0 to 90% @ 30°C, 
                                 0 to 80% @ 40°C, 0 to 70% @ 50°C, noncondensing
Storage Conditions      -40 to 60°C (-40 to 140°F), instrument only Languages 
                                 English, French, German, Italian, Spanish, Portuguese (BR),
                                 Portuguese (PT), Bulgarian, Chinese, Czech, Danish, Dutch,
                                 Finnish, Greek, Hungarian, Japanese, Korean, Polish,
                                 Romanian, Russian, Slovenian, Swedish, Turkish
Interface                    Optional USB
Instrument Enclosure Rating  IP67 (closed lid, battery compartment excluded)
Protection Class         Power Supply: Class II
Certification               CE certified
Sample Required        15 mL (0.3 oz.) Sample Cells
                                60 x 25 mm (2.36 x 1 in.) borosilicate glass with screw cap
Dimensions                22.9 x 10.7 x 7.7 cm (9.0 x4.2 x 3.0 in.)
Weight                      527 g (1.16 lb) without batteries
                                618 g (1.36 lb) with four AA alkaline batteries